Statim G4 Error Code: CYCLE FAULT 1

Statim G4 Error Code: CYCLE FAULT 1:

The cassette temperature failed to reach 95°C within a time-out period

 

Display/Symptom Things To Check Solution
Cycle Fault 1

During Sterilization Mode, chamber temperature reached 6°F above the designated sterilization temperature

 

  1. An extremely large steam leak (Statim 5000). Replace cassette seal, lid or tray
  2. An extremely large load (Statim 5000)
  3. Boiler does not heat up, check thermal fuse by performing a Continuity Test using an ohmmeter. Check for continuity between J13 & lower terminal on boiler. If reading is 1, thermal fuse is defective and needs to be replaced. If reading is 0, then thermal fuse is good. If the fuse is defective, replace and check water pump using Pump Tester Bottle
  4. Check resistance reading between J13 & J14 on board. Reading should be approximately 11 Ω (ohms)
  5. Check for line voltage at terminals J13 & J14 during warm up. If the reading is good proceed to next step. If bad, PCB (Printed Circuit Board) is defective
  6. Check for line voltage at boiler terminals
  1. Replace Cassette Seal Part #SCS029-247
  2. Reduce Load (Remove a few items from the cassette)
  3. Replace Thermal Fuse: Statim 2000 G4 Part #SCF006-9029 or Statim 5000 G4 Part # SCF006-9061
  4. Contact us
  5. Contact us
  6. Contact us

 




This product was added to our catalog on Wednesday 19 September, 2018.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.250 - Number of Queries: 162 - Query Time: 0.118751887146
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale