ERROR CODE C382: FILL MODE

This Troubleshooting Guide is applicable for the following Midmark Autoclave Model Numbers: M9-020 thru -022, M9D-020 thru -022, M11-020 thru -022, M11D-020 thru -022
 

Notice Green LCD Display

 


If Your Midmark has a Green LCD Display as shown above, then this is the right guide for you

 




Midmark M9 & M11 Error C382 (FILL)





Display/Symptom Cause What to Check Solution
C382 FILL MODE
DOOR OPEN
ITEMS NOT STERILE
CLOSE DOOR
Door switch contacts opened during Fill Mode Check with operator to determine if door was opened during Fill Mode Explain proper operation
Close door
Cycle will continue where it left off
If the door is hard to close, inspect door & dam gaskets for proper installation

Be sure trays are pushed all the way in
Install door / dam gaskets properly

Remove any obstructions that may prevent the trays from sliding in completely
Perform: Door Switch Continuity Test Follow instructions in test procedures

Click Here for the Midmark M11 Door Switch Part #MIS095-4035

Click Here for the Midmark M9 Door Switch Part #MIS0095-4036