This test consists of measuring the amount of resistance in the element. To do this, first, turn the autoclave off and unplug it from the wall


WARNING: CURRENT MUST NOT BE PRESENT WHEN MEASURING RESISTANCE.

Set your volt/ohm meter on 200 ohms (200 NOT 200k). The ohms range is located within the Omega Ω (look for this symbol) section of your meter. It is also the area with "k" ranges (ie: 2000k,. 200k, 20K and then 2000, 200).

Place a probe from the volt/ohm meter on each terminal of the 2 terminals (# (8) in figure 4.6 below) on the element (it really doesn't matter which color probe goes where).

For the heating element to be good, it must fall within the resistance range in this table (between 10.80 and 13.20 ohms

If your reading is outside of this range, then it is defective and needs to be replaced.

Model Volts Watts Amps Resistance Resistance Range
Midmark M7 118v 1130 Watts 9.58 Amps ~12.0 ohms 10.80 13.20ohms


Midmark M7 Heating Element


This product was added to our catalog on Sunday 12 April, 2009.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.408 - Number of Queries: 190 - Query Time: 0.156360197891
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale