ERROR CODE C384: STERILIZE MODE




Midmark M9 & M11 C384: STERILIZE MODE





Display/Symptom Cause What to Check Solution
C384: STERILIZE MODE

DOOR OPEN
ITEMS NOT STERILE
UNPLUG / REPLUG UNIT
Door switch contacts opened during Sterilize Mode Perform: Door Switch Continuity Test Follow instructions in test procedures

Click Here for the Midmark M9 Door Switch Part #MIS095-4036

Click Here for the Midmark M911Door Switch Part #MIS095-4035

allclaveparts.com widget logo