Testing the Door Motor
Testing the Door Motor
How To Test The Midmark Autoclave Door Motor System
Testing the Door Motor Assembly on the Midmark Autoclaves is simple, once you know how. Simply Follow These Step-By-Step Directions
Before Starting, you will need to remove the right hand side panel (as you are facing the autoclave) by removing the two screws on top and then lifting out the panel
Procedure | Preview |
Step 1: Remove the Two Wires From The Motor | ![]() |
Door Motor With Wires Removed | ![]() |
Step 2: Set meter to 20K ohms Ω | ![]() |
Step 3: Place Meter Probes On Terminals | ![]() |
Step 4: Look At Readings on Your Meter
The Acceptable Range for 120V Autoclaves is 6620 Plus or minus 20% The Acceptable Range for 230V Autoclaves is 12,200 Plus or minus 20% |
|
If reading is out of acceptable range...Replace door motor. But, If reading is within acceptable range... Perform PC Board Relay Test | Click Here For The M9 Door Motor Assembly
Click Here for the M11 Door Motor Assembly |