Part #MIG157K-8155
Midmark Part #002-0361-01
Midmark M9 Door & Dam Gasket Kit For Units With Metal Flap
Model This Part Fits: Midmark M9 Autoclaves With Shelf Above Dam Gasket
So, where's the wire?
The original Midmark Door Gaskets require a wire to hold them in place - ours do not....Because they are not needed with our New & Improved Midmark M9 Door Gaskets!
When you order one of ours, you can throw the wire away (using the wire on one of our door gaskets will cause a leak). All you need is our Door Gasket
Includes
Get Everything You Need For Annual Preventative Maintenance With Our Midmark M9 PM Kit For Doors With The Metal Flap Part #MIK158-6029
If You See Steam Leaking Around The Door
Then this Door & Dam Gasket Kit Is What You Need To Fix It
The Door & Dam gaskets should be replaced immediately if leakage is noticed. The leakage will usually result in an aborted cycle and a Pressure Related error code
When this happens, your instruments Will Not be sterilized
Click Here For Instructions On How To Install Your New Door Gasket
Click Here For Instructions On How To Install Your New Dam Gasket It is recommended you replace the Door & Dam Gasket every 12 months