ERROR CODE C232: FILL MODE




Midmark M9 & M11 Error C232 (FILL MODE)





Display/Symptom Cause What to Check Solution
C232: FILL MODE
WATER LOW
ITEMS NOT STERILIE
PUSH STOP TO RESTART

Water level sensor did not detect water within the 5 minute time limit
Reservoir water level is low. Check water level in reservoir Add water to the reservoir until level is in the correct range

Press STOP button, then initiate a new cycle

Press STOP button, then initiate a new cycle
Chamber filter is dirty or clogged Inspect the Chamber Filter Clean or Replace the Chamber Filter

Click Here for the Replacement Chamber Filter For The Midmark M9 Autoclave (Part #MIF062-1195)

Click Here for the Replacement Chamber Filter For The Midmark M11 Autoclave (Part #MIF062-798)
Water Level Sensor is dirty or defective Inspect Water Level Sensor Clean or Replace the Water Level Sensor

Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-1217)
Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-815)
Loose / damaged wire connection(s) Check wire connections at these locations:
  • J8 on main PC board
  • Water level sensor terminal
Secure, repair or replace loose or damaged connections
Chamber is dirty or corroded Inspect the bottom of the chamber Clean the bottom of the Chamber with Mid-Clean™ and distilled water Part #MIC143-6136
Fill Valve or PC Board malfunctioning Inspect Fill Valve and related tubing for blockages Clean, repair or replace Fill Valve
Inspect two wires connecting the Fill Valve and PC board (W3 & W4 on PC board) for damage and/or loose connections Secure any loose connections
With the sterilizer in the Fill Mode...Perform: Fill Valve Testing Follow instructions in test procedures
Air valve or PC board malfunctioning With the sterilizer in the Fill Mode, Perform Air Valve Testing Follow instructions in test procedures



This product was added to our catalog on Saturday 13 September, 2014.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.368 - Number of Queries: 201 - Query Time: 0.212357431107
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale