Error Code E005

Midmark M9 & M11 Error Code E005


Note: If you started receiving this error code E005 after replacing the Bellows or Steam Trap, check your bellows to make sure you did not install it backwards. The printing and direction arrow should point upwards toward the Water Reservoir

Display/Symptom Cause What to Check
E005 - Excessive Pressure
(Pressure Exceeded 35psi)
Sterilizer is overloaded Reduce Load
Bellows (or Steam Trap) is closing prematurely trapping air in the chamber Run a cycle while watching the Bellows inlet pipe in the Water Reservoir and listening to the air rushing through the Bellows. As the unit heats the water, you should hear the air coming through the Bellows get louder & louder

You will hear the tone of the rushing air going through the bellows start changing at about 202°F. and should stop at about 212-215°F. If the Bellows is closing at say...209 degrees, then it is trapping cold air inside the chamber and is the cause for the over pressurization. The Bellows or Steam Trap will need to be replaced
Temperature probe is dirty or malfunctioning Clean Temperature Probe. If no change, unplug the Temperature Probe From The PC Board and Try another Cycle. If the error code goes away, then the Temperature Probe is defective and needs to be replaced
Sterilizer potentiometers are out of calibration Recalibrate the Autoclave with our Midmark Calibration Kit
Heating Element is not Turning On & Off Check to make sure your heater is turning on and off. You can check this by watching the display once you enter the count down cycle. You should heard the heater relay turn off, a loud click, and see the temperature drop a few degrees and then hear the heater relay turn on, another click and see the temperature rise. This cycle should happen every few minutes during the sterilization cycle
You have a blockage in one of you copper lines going to the chamber Clear the Copper Tubing of any blockages



This product was added to our catalog on Monday 04 May, 2015.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.279 - Number of Queries: 216 - Query Time: 0.114654923172
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale