OCM PCT007-055 Solid State Controller Installation Instructions

PCT007 & PCT169
SOLID STATE CONTROLLER
WIRING INSTRUCTIONS AND DIAGRAMS


BEFORE REPLACING THERMISTOR OR SOLID STATE HEAT CONTROLLER, PLEASE CHECK THE FOLLOWING:
  • Make sure there is sufficient water in chamber
  • Check fill tube and valves for leaks
  • Check heaters and all connections
  • Check Overtemperature Thermostat (Part #PCT042). Reset or replace if necessary
When replacing the solid state controller or thermistor, note Serial Number of autoclave and install per corresponding wiring diagram to the right

Solid state controller is supplied with choke and M.O.V. mounted P.C. board.

For Machines Using Both Steam and Dry Heat Sterilizing Modes
  1. Mount controller and rewire per wiring diagram
  2. Press in interlock stop button and turn temperature control fully clockwise
  3. Each installation REQUIRES individual adjustment to match the individual machine as follows:

    • If steam and dry heat are required (dry heat door must be installed) bring temperature up to 350°F by adjusting rheostat on P.C. board. (Note: The "Heat-On" lamp will flash before final temperature is reached. Allow 10-15 minutes for temperature to stabilize before each adjustment.)
    • Turn machine off and allow to cool
    • Turn temperature control counter-clockwise until interlock button pops out. Remove knob from shaft.
    • Operate machine through steam cycle. Adjust front rheostat to stabilize temperature at 270°F. (Note: The "Heat-On" lamp will flash approximately 10° before temperature stabilizes. Allow 10-15 minutes for stabilization after each adjustment.)
    • Replace temperature control knob. Make sure flange on knob is against interlock stop button
For Machines Using Steam Sterilization Only
  1. Mount controller and rewire per diagram
  2. Operate sterilizer through steam cycle with front control rheostat turned clockwise to interlock stop button (maximum setting for steam sterilization)
  3. Each installation REQUIRES individual adjustment to match the individual machine as follows:


    • Adjust rheostat on side of printed circuit board until sterilizer temperature stabilizes at 270°F. (Note: Approximately 10°F before temperature stabilizes the "Heat- On" lamp will start flashing. Allow 10-15 minutes for stabilization of temperature after each front control rheostat adjustment.)
    • Run machine for several cycles to verify setting. Make any necessary readjustments.


This product was added to our catalog on Sunday 24 January, 2010.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.302 - Number of Queries: 194 - Query Time: 0.135203772659
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale