(image for) Midmark M11D Ultraclave TEMPERATURE SENSOR

Midmark M11D Ultraclave TEMPERATURE SENSOR

Model: MIS121-7469

$119.84

Description

High Quality Midmark M11D Temperature Sensor
Part #MIS121-7469
Midmark Part #015-1680-00

TEMPERATURE SENSOR

  • 1/4 MPT
  • Stainless steel housing
  • Range: ~1.07 to 1.1O @ 25°C (Positive Temperature Coefficient)
  • Includes 18-1/2" long wire with connector
  • See Notes below for Serial Numbers this part fits

Fits: Rear Chamber Port

Model This Part Fits: Midmark M11D Autoclaves

Notes: CHECK SERIAL NUMBERS HERE:

  • Fits M11D-020 thru -022 with Serial Number prefixes RY, RZ & V


The M11D Temperature Sensor is located in the back of the chamber.

 

 

How To Test the Temperature Sensor

Warning: Power Must Not Be Present When Taking Resistance Readings. Disconnect power to the unit before performing this test.

Note: Let the unit cool down before testing. The Temperature Sensor must be tested at room temperature (25°C or 77°F)

 

 

 

 

  • Disconnect the Temperature Sensor wire harness from the J12 connector on the PC Board
  • Set the volt/ohm meter to 2k ohms
  • Place the multimeter probes on the red and white wires of the Temperaure Sensor wire harness

Your reading on your meter should be between 1.07 and 1.1 ohms. If it is outside of this range, replacement Of your M11 Temperature Probe is required

 

 

 

 

Midmark M11 Temperature Sensor

 

 

 

 

What The M11D Temperature Sensor Does

This sensor monitors the chamber temperature during the Heat-up and Sterilization Modes, and send this information back to the PC Board.

Based on this feed back, the PC Board turns the heating element On and Off as needed

Service Tip: Residue can build up on the Probe of the sensor causing false temperature readings.

If a temperature related malfunction occurs, clean the probe with an abrasive pad (Scotch Brite White Or Pink)

 

 

 

 

Temperature Sensor Function

 

 

During The Fill Mode The Temperature Is Not Monitored
During The Heat Up & Sterilization Modes The Temperature Sensor Continually Monitors The Chamber Temperature & Transmits This Information To The PC Board

The PC Board Turns The Heating Element ON/OFF Based on The Readings From The Temperature Sensor
During The Vent Mode The Temperature Sensor Continually Monitors The Chamber Temperature & Sends This Information To The PC Board
During The Drying Mode The Temperature Sensor Continually Monitors The Chamber Temperature & Transmits This Information To The PC Board

If The Temperature Sensor Exceeds 240°F (115°C), The Board Stops The Current Flow To The Heating Element Until The Temperature Drops
Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods
allclaveparts.com widget logo