(image for) Midmark M9 FIELD SERVICE CALIBRATION SMART KIT

Midmark M9 FIELD SERVICE CALIBRATION SMART KIT

Model: MIK074-1206

$398.36

Description

Midmark M9 High Quality Guaranteed Replacement Parts

Used By Professional Autoclave Repair Technicians World Wide

EASY TO USE FIELD SERVICE CALIBRATION SMART KIT


Part #MIK074-1206

FIELD SERVICE CALIBRATION SMART KIT

  • Use to check and adjust both Temperature & Pressure
  • For use with older style units

Includes:

Model(s) This Part Fits: This One Kit Can Be Used To Service Both The Midmark M9 & The Midmark M11 Autoclaves!

Including

  • M9 Autoclaves With Serial Number Prefixes: CZ, DA, DB, DX, DY, FD & OM)
  • M9D Autoclaves With Serial Number Prefixes: FF & FG)
  • M11 Autoclaves With Serial Number Prefixes: ES, ET, FP, FR & GB
  • M11D Autoclaves With Serial Number Prefixes: GC & GD)

Easy To Follow Step-By-Step Directions Included

This kit is required when you get an Error Code "009" and recalibration is required Click Here For Instructions

If you have a technician recalibrate it for you, this is the kit he will use to perform the procedure. The average cost for this service is $250.00. So you will be money ahead the very first time you use it.
 

Error codeSymptomProbable CauseCheckProcedure
Error Code 9M9/M11 - Door opens at end of cycle.

M9D/M11D - Door is manually opened at end of cycle
Zero pressure potentiometer out of calibrationCheck calibration of zero pressure potentiometerAdjust zero pressure potentiometer.

Click Here For Instructions
 M9/M11 - Door does not open at end of cycle.Latch lever & pulse solenoid plunger mechanism malfunctioningPerform M9/M11 Door Opening Test.

Click Here For Instructions
Remove any debris or buildup from latch lever & pulse solenoid mechanism.
  Door pins and latch bracket not functioning properlyCheck for burrs on door pins and latch bracket in doorRemove burrs; then clean and lube with lithium based grease.
 Zero pressure potentiometer out of calibrationCheck calibration of zero pressure potentiometer.Adjust zero pressure potentiometer.

Click Here For Instructions
 
 Door is closed when error is detectedDoor was closed during Dry cycle Inform operator not to shut door during Dry cycle.

 

Pays For Itself Over & Over Again 
 

 

"Midmark M9 Autoclave Repair Made Simpleā„¢"

 

 