Description

Used By Professional Autoclave Repair Technicians World Wide
EASY TO USE FIELD SERVICE CALIBRATION SMART KIT
Part #MIK074-6983
FIELD SERVICE CALIBRATION SMART KIT
- Use to check and adjust Pressure & Temperature
- Includes: See Notes
- All parts in kit are sold separately
Model(s) This Part Fits: This One Kit Can Be Used To Service Both The Midmark M9 & The Midmark M11 Series Autoclaves!
- Notes - CHECK SERIAL NUMBERS HERE: Fits M9-001 thru -019 with Serial # prefixes CZ, DA, DB, DX, DY, FD & OM. Fits M9D-001 thru -019 with Serial # prefixes FF & FG. Fits M11-001 thru -019 with Serial # prefixes ES, ET, FP, FR & GB. Fits M11D-001 thru -019 with Serial # prefixes GC & GD.
Easy To Follow Step-By-Step Directions Included
This kit is required when you get an Error Code "009" and recalibration is required. If you have a technician recalibrate it for you, this is the kit he will use to perform the procedure. The average cost for this service is $250.00. So you will be money ahead the very first time you use it.
Notes: Includes:
| Error code | Symptom | Probable Cause | Check | Procedure |
| Error Code 9 | M9/M11 - Door opens at end of cycle.
M9D/M11D - Door is manually opened at end of cycle | Zero pressure potentiometer out of calibration | Check calibration of zero pressure potentiometer | Adjust zero pressure potentiometer.
Click Here For Instructions |
| | M9/M11 - Door does not open at end of cycle. | Latch lever & pulse solenoid plunger mechanism malfunctioning | Perform M9/M11 Door Opening Test.
Click Here For Instructions | Remove any debris or buildup from latch lever & pulse solenoid mechanism. |
| | | Door pins and latch bracket not functioning properly | Check for burrs on door pins and latch bracket in door | Remove burrs; then clean and lube with lithium based grease. |
| | Zero pressure potentiometer out of calibration | Check calibration of zero pressure potentiometer. | Adjust zero pressure potentiometer.
Click Here For Instructions | |
| | Door is closed when error is detected | Door was closed during Dry cycle | | Inform operator not to shut door during Dry cycle. |
Pays For Itself Over & Over Again
"Midmark M9 Autoclave Repair Made Simpleā¢"