For the simplist & easiest installation possible, we highly recommend you
do not remove the old bellows assembly until you have the new one (PCB237-5875) in hand
How To Replace the Bellows (Pressure Switch)On The Pelton Crane Magnaclave
Here's How To Remove the Bellows Assembly (Pressure Switch)
- Remove the pressure stat assembly from the machine
- Remove the micro-switch
- Rotate the control shaft completely, allowing the spring to fall free of the actuator and Remove Spring
- Remove the E ring from the bellows
- Remove the hex nut from the bellows and then remove the defective bellows from the assembly
Here's How To Install The New Bellows Assembly (Pressure Switch)
- Working the repair in reverse, install the new bellows
- Using your fingers on one hand to depress the actuator and compress the bellows, push the spring back onto
the stainless steel bushing with your other hand
- Rotate the control shaft and verify proper movement of the spring against the actuator
- Reinstall the micro-switch
- Reinstall the pressure stat assembly in the machine
- Before installing the knob back on the pressure stat, run a sterilization cycle to orientate the knob with the
temperature that is registered on the gauge