Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Midmark Troubleshooting
› ERROR CODE C326: DOOR MODE
ERROR CODE C326: DOOR MODE
Midmark M9 & M11 Error C326: DOOR MODE
Display/Symptom
Cause
What to Check
Solution
C326: DOOR MODE
DOOR CLOSED
OPEN DOOR
Door switch indicates that the door remains closed one second after door motor has attempted to open door.
Door / dam gasket(s) sticking
Trays not pushed in properly
Inspect Door and Dam gaskets for proper installation and any signs of sticking
Be sure trays slide in properly
Install door / dam gaskets properly
Remove any obstructions that may prevent the trays from sliding in completely
Click Here
For the Midmark M9 Door & Dam Gasket Kit Part # MIG027K-3380
Click Here
For the Midmark M9 Door & Dam Gasket Kit ACPMM11-3381
Door spring missing and/or damaged
Inspect door spring for damage
(If the door motor releases the latch, but the door remains closed: replace the door spring)
Replace door spring if necessary
Click Here
for the Replacement Door Spring For The Midmark M9 Autoclave (Part #MIS076-1218)
Click Here
for the Replacement Door Spring For The Midmark M11 Autoclave (Part #MIS076-816)
Door motor system / latch "hanging up"
Remove the door cover, and RH side panel
With the door closed, run the "Door Open Test" (
Click Here
For Instructions). Watch for any mechanical issues that may prevent the door from opening
Refer to: Service Diagnostics (I/O Test) (
Click Here
for Instructions)
If mechanical issues are apparent, Adjust / replace door motor system as necessary
Door pins / latch binding
With door open, door handle should move Up & Down freely
Inspect latch mechanism for wear / damage
Clean door pins & latch mechanism
Replace worn / damaged components
Door switch or PC board malfunctioning
Perform: Door Switch Testing
Follow instructions in test procedures
Residual pressure in chamber
Check chamber pressure when door motor is energized
If pressure is above 0.72 psi (5 kPa), check for blockage in the pressure sensor tubing between the PC board and the chamber
In addition, inspect the filter located inside the chamber to be sure it is not clogged
Clear any restriction that could prevent chamber pressure from reaching the pressure sensor on PC board