Midmark M9 & M11 Error Code E009
Door Opens At End Of Cycle
Display/Symptom |
Cause |
What to Check |
Solution |
Error code 9
[Pressure exceeded 0.9 psi (6.2 kPa) during Dry cycle]
Door Automatically Opens At End Of Cycle
(Door Manually Opened on M9D & M11D Autoclaves) |
- Zero Pressure Potentiometer out of Calibration
|
- Check Calibration of Zero Pressure Potentiometer
|
- Adjust Zero Pressure Potentiometer
|