Base Inspection Cover Removal/Installation

 

 

Another AllClaveParts Exclusive:  The Most Comprehensive Tuttnauer Autoclave Troubleshooting & Maintenance Guides To Be Found Anywhere - At Any Price - Can Only Be Found Here....... And They Are Absolutely Free! 

 

 

Removing the Base Cover Allows you to
Test & Replace the Following Components

  1. The Heating Element Terminals
  2. The High Limit (Overheat) Thermostat, And
  3. Check For Leaks Where the Heating Element Rods Pass Through The Chamber
A. Removal
  1. Disconnect power cord from outlet
  2. Drain water from condensing tank
  3. Turn sterilizer onto its back
  4. Remove two screws (1) and base inspection cover (2) from base 3)



B. Installation
  1. Install base inspection cover (2) on base (3) and secure using two screws (1)
  2. Turn sterilizer upright
  3. Refill condensing tank with distilled water