Shopping
Autoclave Maintenance
Autoclave Parts
Autoclave Troubleshooting
Dental Equipment Parts
MEDIVATORS
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
Account Info
Log In
Create Account
Autoclave Maintenance
Tuttnauer Maintenance Guides
Statim Maintenance Guides
Autoclave Parts
Scican Statim
Midmark ® Ritter
Tuttnauer ®
Pelton Crane
Barnstead Sterilizer
Kavoklave
Midmark Ritter
NAPCO
AMSCO
PRESTIGE 2100
Ritter - Castle
Spectroline
S.S. White
Vernitron
Autoclave Cleaning Products
MDT Harvey Mediclaves
Autoclave Repair Supplies
Autoclave Repair Special Tools
Autoclave Troubleshooting
Prestige 2100
Statim G4 Troubleshooting
tuttnauer temp
STATIM DETAILED TROUBLESHOOTING
VernitronTroubleshooting
NAPCO Troubleshooting
How To Test Heating Elements
Midmark Troubleshooting
Tuttnauer Troubleshooting
STATIM TROUBLESHOOTING
Documents & Guides
Pelton Crane Troubleshooting
Troubleshooting Guides
AMSCO 3000 BASIC TROUBLESHOOTING
AMSCO 900 & 901
Delta XL Troubleshooting
Castle Table Top Troubleshooting
CHEMICLAVE 5000 TROUBLESHOOTING
Kavoklave
Steam Trap Troubleshooting
Dental Equipment Parts
Dental Chair Parts
Dental Handpieces & Parts
DENTAL ULTRASONIC SCALERS
Miscellaneous
Dental Compressor Parts
Dental Delivery Systems
Dental Film & X-Ray Processors
Dental Light Parts
Dental Vacuum Parts
Dental Water Control Parts
MEDIVATORS
Medivator Advantage Plus
Medivator DSD-201
Medivator CER Optima
Medivator DSD-91E
Medivator DSD Edge
Medivator SSD-102
More
Videos and Animations
EXAM TABLES & CHAIRS
Case Studies
Free Technical Support
Autopsy Saw Parts
Infant Incubators & Warmers
Lab Equipment
Ventilators
Orthopedics & Physical Therapy
Patient Cables & Leadwires
ASPIRATORS AND SUCTION PUMPS
Maintenance Guides
Operational Manuals
Maintenance Tips
Parts Installation Guides
About Us
Blog
Home
›
Autoclave Troubleshooting
›
Pelton Crane Troubleshooting
›
VALIDATOR TROUBLESHOOTING GUIDE
› Validator TroubleShooting Guide
Validator TroubleShooting Guide
Troubleshooting The Pelton Crane Validator
Your Guide To Maintaining & Repairing The Pelton Crane Validator 8 & 10 Sterilizer
Troubleshooting
Symptom
Analyze
Remedy
Mineral Build Up Inside Chamber
Type Of Water That Is Being Used
Clean Inside Of Chamber Thoroughly - Refill Using Distilled Water Only
Sterilizer Takes Over 15-20 Minutes (From a Warm Start) To Reach Set Sterilization Temperature
Chamber Load
Place Fewer Packages Inside Chamber
-
Bellows
Replace If Necessary
Validator 8 Bellows
Validator Plus 8 Bellows
Validator 10 Bellows
Validator Plus 10 Bellows
-
Line Voltage
If Line Voltage Is Too Low Plug Into Outlet With Proper Voltage - Contact Qualified Electrician.
-
Solenoid Valves
Refer To
Panel 5
(Water Reservoir Leak Identification To Identify LeaksClean Inside Of Chamber Thoroughly
Displayed Temperature Stays Considerably Below Required Temperature For Sterilization Cycle
Ribbon Cable
If Necessary Replace Ribbon Cable:
Validator 8 Temperature/Pressure Cable
Validator 10 Temperature/Pressure Cable
-
Bellows
Replace If Necessary
Validator 8 Bellows
Validator Plus 8 Bellows
Validator 10 Bellows
Validator Plus 10 Bellows
-
Dump Valve. Refer To
Panel 5
(Water Reservoir Leak Indentification)
If Necessary Clean Dump Valve And/Or Replace Valve Kit
Validator 8 DUMP PLUNGER KIT
Validator 10 DUMP PLUNGER KIT
See Panel 8
Wet Packs After Drying Cycle Is Complete Or Wet Packs Between Layers
Chamber Load
Break Sterization Load Into Fewer Or Smaller Packs/Wraps For Proper Steam Sterilization.Crack Door Open - 1" During Drying Cyle.
-
Fill-Line Filter Condition & Position On Chamber Floor
Replace Fill-Line Filter And/Or Adjust Filter To Center Bottom Of Chamber
Validator 8 FILTER - FILL LINE
Validator 10 FILTER - FILL LINE
Validator Plus 8 FILTER - FILL LINE
Validator Plus 10 FILTER - FILL LINE
Water In Bottom Of Sterlizer When Unit Is Not In Fill Cycle
Fill Valve
Clean Fill Valve And/Or Replace Valve Kit
Validator 8 FILL PLUNGER KIT
Validator 10 FILL PLUNGER KIT
See Panel 8
Water Dripping From Door When Door Is Opened After Cycle
Spacer (Door)
Check Spacer For Shrinkage Or Damage. Replace If Necessary
Validator 8 SPACER (DOOR
)
Validator 10 SPACER (DOOR)
Validator Plus 8 SPACER (DOOR)
Validator Plus 10 SPACER (DOOR)
When Unit Vents, Water Comes Out Of Fill Port On Top Of Case
Water Level Inside Reservoir
Drain Water To The Bottom Holes Of Fill Cup
Sterilizer Fails Several Self-Tests, But Still Operates
Cable Connections
Make Sure Cable Connectors At Front Panel And Power Board Are Good
Sterilizer Does Not Turn On
Cable And On/Off Switch (Power LED On Power Supply Board Should Be Illuminated)
Check MicroProcessor Board On/Off Switch And Cable With Ohm Meter To Verify Continuity* If Switch & Cable Pass Test, Replace Power Supply Board
Sterilizer Does Not Switch Modes
Gap Between Panel Membrane And Switches Mounted On Micro-Processor Board
Check Between MicroProcessor Board And Panel Membrane For Washers Or Materials Causing Incorrect Separation. Replace MicroProcessor Board If Not Correctable.