(image for) Midmark M9D CALIBRATION SMART KIT FOR FIELD SERVICE

Midmark M9D CALIBRATION SMART KIT FOR FIELD SERVICE

Model: MIK074-6983

$398.36

Description

Midmmark M9D High Quality Guaranteed Replacement Parts

 

 

Used By Professional Autoclave Repair Technicians World Wide

EASY TO USE FIELD SERVICE CALIBRATION SMART KIT 

 


Part #MIK074-6983 

FIELD SERVICE CALIBRATION SMART KIT

  • Use to check and adjust Pressure & Temperature
  • Includes: See Notes
  • All parts in kit are sold separately

Model(s) This Part Fits: This One Kit Can Be Used To Service Both The Midmark M9 & The Midmark M11 Series Autoclaves!

  • Notes - CHECK SERIAL NUMBERS HERE: Fits M9-001 thru -019 with Serial # prefixes CZ, DA, DB, DX, DY, FD & OM. Fits M9D-001 thru -019 with Serial # prefixes FF & FG. Fits M11-001 thru -019 with Serial # prefixes ES, ET, FP, FR & GB. Fits M11D-001 thru -019 with Serial # prefixes GC & GD.

Easy To Follow Step-By-Step Directions Included 

This kit is required when you get an Error Code "009" and recalibration is required. If you have a technician recalibrate it for you, this is the kit he will use to perform the procedure. The average cost for this service is $250.00. So you will be money ahead the very first time you use it.

Notes: Includes:

 

Error codeSymptomProbable CauseCheckProcedure
Error Code 9M9/M11 - Door opens at end of cycle.

M9D/M11D - Door is manually opened at end of cycle
Zero pressure potentiometer out of calibrationCheck calibration of zero pressure potentiometerAdjust zero pressure potentiometer.

Click Here For Instructions
 M9/M11 - Door does not open at end of cycle.Latch lever & pulse solenoid plunger mechanism malfunctioningPerform M9/M11 Door Opening Test.

Click Here For Instructions
Remove any debris or buildup from latch lever & pulse solenoid mechanism.
  Door pins and latch bracket not functioning properlyCheck for burrs on door pins and latch bracket in doorRemove burrs; then clean and lube with lithium based grease.
 Zero pressure potentiometer out of calibrationCheck calibration of zero pressure potentiometer.Adjust zero pressure potentiometer.

Click Here For Instructions
 
 Door is closed when error is detectedDoor was closed during Dry cycle Inform operator not to shut door during Dry cycle.

Pays For Itself Over & Over Again 
 

"Midmark M9 Autoclave Repair Made Simpleā„¢"

 

 

Fast Delivery

We usually ship the same day! All packages are shipped w/Delivery Confirmation.

100% Secure Payment

Your transaction is secure thanks to the PayPal payment system: SSL (Secure Socket Layer).

Shopper Award
(image for) AllClaveParts Payment Methods
allclaveparts.com widget logo