How To Remove And Install The M9 Fill Solenoid Valve

Midmark M9 Fill Solenoid Removal / Installation Instructions

NOTE Steps will vary slightly for M9D units. Use this procedure as a guide.

Removal

NOTE Units prior to serial number CZ1110/OM1000 have a manifold assembly which secures the fill solenoid.

Refer to the parts list to remove the fill solenoid on these sterilizers.

  1. Drain water from condensing tank.
  2. Remove back panel ( See Instructions Here ).
  3. Tag and disconnect two wires (1, Figure Below) from terminals (3).
  4. Tag and disconnect two wires (2) from terminals (4).
  5. Disconnect three tubes (5 thru 7) from elbows (8 thru 10).
  6. Remove two screws (11) from bottom of manifold assembly (12).




  7. NOTE: Spacer (12A) is only on newer units.


  8. Remove manifold assembly (12) and spacer (12A) from base (13).
  9. Hold screw (1, Figure 4-13) and loosen nut (2). Turn terminals out of the way.
  10. Remove assembled vent solenoid (3) and elbow (4) from tee (5).
  11. Remove elbow (6) and tee (5) from fill solenoid (7).
  12. Remove elbow (8) from fill solenoid (7)

Installation

  1. Coat threads of tee (5, Figure Below) and elbows (6 and 8) with teflon tape.
  2. Install elbow (8) on fill solenoid (7).
  3. Install elbow (6) and tee (5) on fill solenoid (7).
  4. Loosen nut (9). Turn terminals of fill solenoid (7) so wires will connect easily to fill solenoid. Tighten nut (9).
  5. Coat threads of elbow (4) with teflon tape.
  6. Hold screw (1) and loosen nut (2). Turn terminals out of the way.
  7. Install assembled elbow (4) and vent solenoid (3) on tee (5).
  8. Turn terminals of vent solenoid (3) so wires will connect easily to vent solenoid.
  9. Hold screw (1) and tighten nut (2).


  10. Warning: When installing the vent solenoid and the fill solenoid, the side of the solenoids marked "IN" need to be connected to the pressure lines. Reversing solenoids will cause system failure.

  11. Install spacer (12A, Figure 4-12) and manifold assembly (12) on base (13) and secure using two screws (11).
  12. Connect three tubes (5 thru 7) to elbows (8 thru 10).
  13. Connect two wires (2) to terminals (4).
  14. Connect two wires (1) to terminals (3).
  15. Install back panel (Refer to para 4.5).
  16. Refill condensing tank with distilled water.


This product was added to our catalog on Wednesday 25 February, 2009.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.170 - Number of Queries: 183 - Query Time: 0.0689915332184
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale