ERROR CODE C232: FILL MODE
ERROR CODE C232: FILL MODE
Midmark M9 & M11 Error C232 (FILL MODE)
Display/Symptom | Cause | What to Check | Solution |
C232: FILL MODE
WATER LOW ITEMS NOT STERILIE PUSH STOP TO RESTART Water level sensor did not detect water within the 5 minute time limit |
Reservoir water level is low. | Check water level in reservoir | Add water to the reservoir until level is in the correct range
Press STOP button, then initiate a new cycle Press STOP button, then initiate a new cycle |
Chamber filter is dirty or clogged | Inspect the Chamber Filter | Clean or Replace the Chamber Filter
Click Here for the Replacement Chamber Filter For The Midmark M9 Autoclave (Part #MIF062-1195) Click Here for the Replacement Chamber Filter For The Midmark M11 Autoclave (Part #MIF062-798) |
|
Water Level Sensor is dirty or defective | Inspect Water Level Sensor | Clean or Replace the Water Level Sensor
Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-1217) Click Here for the Midmark M9 Water Level Sensory Assembly (Part #MIS075-815) |
|
Loose / damaged wire connection(s) | Check wire connections at these locations:
|
Secure, repair or replace loose or damaged connections | |
Chamber is dirty or corroded | Inspect the bottom of the chamber | Clean the bottom of the Chamber with Mid-Clean™ and distilled water Part #MIC143-6136 | |
Fill Valve or PC Board malfunctioning | Inspect Fill Valve and related tubing for blockages | Clean, repair or replace Fill Valve | |
Inspect two wires connecting the Fill Valve and PC board (W3 & W4 on PC board) for damage and/or loose connections | Secure any loose connections | ||
With the sterilizer in the Fill Mode...Perform: Fill Valve Testing | Follow instructions in test procedures | ||
Air valve or PC board malfunctioning | With the sterilizer in the Fill Mode, Perform Air Valve Testing | Follow instructions in test procedures |