Check Cassette 1

Check Cassette #1



Note: Before starting, verify that the Statim says Select A Program when turned on. All leaks should be corrected before proceeding with troubleshooting

What Caused the Error Message "Check Cassette #1"

This error is generated when the cassette temperature failed to reach 95°C in three minutes. This means the boiler is not heating up - Generally caused by no power to the boiler

What do you do when this happens?

  1. Check the Thermal Fuse. Using an ohmmeter, check for continuity between J1-3 on the board & the lower terminal on the boiler. If you read less than 1-ohm the thermal fuse is good. If thermal fuse checks good proceed to step 4, if bad replace the Thermal Fuse with and check unit for proper operation.
  2. Test Water Pump using the Water Pump Tester
  3. Check resistance reading between J1-3 & J1-4 pins on the board. Your meter should read approx. 11ohms
  4. Check for line voltage at terminals J1-3 & J1-4 on the board during warm up. If the reading is good proceed to step 5; if bad PCB (Printed Circuit Board) is defective and needs to be replaced
  5. Check for line voltage at boiler terminals.
  6. Check for large steam leak coming from Cassette Bay & Around Cassette. If leaking is present:
    • Remove Cassette Seal
    • Clean seal groove
    • Lubricate seal with soap
    If leakage continues, replace Cassette Seal
  7. Over-loaded Cassette Tray - Reduce load and try another cycle
  8. Inspect and clean steam ports on probe bracket (see Picture Below
Procedure Part Description Part Number For Statim 2000 Part Number For Statim 5000
1 Thermal Fuse Part #SCF006-377 SCF006-517
2 Water Pump Tester SCK024-384 SCK024-522
5 Liquid Soap RPS287-242 RPS287-507
5 Cassette Seal SCS001-241 SCS029-247


Ports


This product was added to our catalog on Wednesday 18 March, 2015.



Customer Reviews
Solution Graphics
Pay By Credit Or Debt Card
-Pay By Credit Card-
Proceed To Checkout And Enter Your Credit Card Information In Step 2

Pay By Credit Or Debit Card
To Order By Phone Call
636-390-2792
Mon thru Fri 8:00am To 4:30 pm Central Time

Testimonials

Don't Take Our Word For It.....


Customer Reviews


AllClaveParts.com - Real Help With Your Autoclave Repair When You Need It

"My Statim 2000 stopped working.

Benco dental came down and said the thermal fuse and computer board was bad

Real Problems Solved By Dr Gorenburg With AllClaveParts help

Wally, thank you and your company for providing great customer service and all the help I needed to fix my autoclave.
Clearview Medical.com

Dr Haruki & Son

My tech told me my STATIM 2000 repair could either be about $350 (if it was only the thermal fuse) or up to $1500 if it was another problem.

Craig Haruki DDS
Lihue, HI


Dr Whitehouse Loves AllClaveParts.com
"It was a pleasure working with Wally who saved me thousands over what I would have had to pay when the Patterson tech......"

Dr Joseph Whitehouse
Castro Valley, CA
Create A Smile


Dr Steve Says Midmark Told Him To Throw It Away, But AllClaveParts Showed Him How To Fix It Well, that same autoclave is once again working like a champ thanks to allclaveparts.com.

Steve Douglas, DDS
Indianapolis, IN
stevedouglasddspc.com


Parse Time: 0.240 - Number of Queries: 191 - Query Time: 0.110965414551
Content

Check Out Our Specials


A Best Seller
Click Here To Check Out Everything On Sale