(image for) Midmark M11 TEMPERATURE PROBE ASSEMBLY M11-001 thru -019

Midmark M11 TEMPERATURE PROBE ASSEMBLY M11-001 thru -019

Model: MIP050-811

$152.93

Description

Guaranteed High Quality Midmark M11 Autoclave Repair Parts

Part #MIP050-811
Midmark Part #002-0357-00

TEMPERATURE PROBE ASSEMBLY

  • 1/8 MPT
  • See Notes below for Serial #'s this part fits
  • See Instructions Below For Testing The Temperature Probe

Fits: Chamber

Model This Part Fits: Midmark M11 Autoclave

Notes: This part fits the Midmark M11-001 thru -019 with Serial # prefixes ES, ET, FP, FR & GB

 

Here is where to Find The Temperature Probe Assembly
Inside The Midmark M11 Autoclave Chamber
Looking Inside The Midmark M11 Chamber
Click on Image For Larger Picture

 

Symptoms Of A Shorted Temperature Probe

 

A shorted Temperature Probe can cause the following symptoms:

• The sterilizer will not "boot up" and will sound a steady beep.

• The display will show a steady temperature reading of ~30˚ C  without changing (this is not normal). The sterilizer will then continue into the sterilize mode and will ultimately overheat, causing an error code to be displayed.

How To Test The Midmark M11 Temperature Probe

 

  • Disconnect the Temperature Probe from the Control PC Board and measure the resistance at room temperature (~72 Degrees F)
  •  The resistance between the BLACK and WHITE wires should read between 3.0 Meg Ohms and 6.0 Meg Ohms. If resistance readings are not met, replace the Temperature Probe.
  •  The resistance between the BLACK and RED wires should read between 3.0 Meg Ohms and 6.0 Meg Ohms. If resistance readings are not met, replace the Temperature Probe.
  •  The resistance between the WHITE and RED wires should read between 70 K Ohms and 80 K Ohms. If resistance readings are not met, replace the Temperature Probe. Connect the Temperature Probe to the Control PC Board and power up the sterilizer.
  •  Verify presence of 5 VDC on the temperature probe between the BLACK and WHITE (ground) wires.
  •  If voltage is not present check ground connections on the PC Board. 
  • If voltage reading is not 5 VDC, replace Control PC Board (Part #MIB131-4220)